6-23
Trace Data Transfers
Internal Measurement Arrays
Error Coefficient Arrays
The error coefficient arrays are either default values or are created
during a measurement calibration. These are used whenever correction
is on. They contain complex number pairs, are accessible via GPIB, and
are referenced as
CH[1|2]SCORR1
,
CH[1|2]SCORR2
,
CH[1|2]SCORR3
and
CH[1|2]SCORR4
.
Table 6-5
Error Coefficient Arrays
NOTE
These arrays do not apply to Broadband External measurements.
Selected Measurement
Error Coefficient Arrays
Transmission (B/R) Response
CH[1|2]SCORR1
Tracking
Transmission (B/R) Response & Isolation
CH[1|2]SCORR1
Tracking
CH[1|2]SCORR2
Isolation Term
Transmission (B/R) Enhanced Response
CH[1|2]SCORR1
Directivity
CH[1|2]SCORR2
Source Match
CH[1|2]SCORR3
Reflection Tracking
CH[1|2]SCORR4
Transmission Tracking
Reflection (A/R)
CH[1|2]SCORR1
Directivity
CH[1|2]SCORR2
Source Match
CH[1|2]SCORR3
Tracking
Broadband Internal
CH[1|2]SCORR1
R* Response
Summary of Contents for 8712ES
Page 11: ...1 1 1 Introduction to GPIB Programming ...
Page 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Page 36: ...3 1 3 Passing Control ...
Page 39: ...4 1 4 Data Types and Encoding ...
Page 46: ...5 1 5 Using Status Registers ...
Page 71: ...6 1 6 Trace Data Transfers ...
Page 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Page 99: ...7 1 7 Using Graphics ...
Page 105: ...8 1 8 Front Panel Keycodes ...
Page 111: ...9 1 9 Introduction to SCPI ...
Page 129: ...10 1 10 Menu Map with SCPI Commands ...
Page 268: ...12 1 12 SCPI Conformance Information ...
Page 290: ...13 1 13 SCPI Error Messages ...