![Agilent Technologies 8712ES Programmer'S Manual Download Page 150](http://html.mh-extra.com/html/agilent-technologies/8712es/8712es_programmers-manual_2867956150.webp)
10-22
Programmer’s Guide
Menu Map with SCPI Commands
Menu Map for 8712ET/ES and 8714ET/ES
Table 10-7
CAL
, Functions, 8712ET/ES and 8714ET/ES
1. Pressing this key begins a guided calibration procedure using three standards.
When the calibration is complete, the command
*WAI;:SENS[1|2]:CORR:COLL:
SAVE;*WAI
is executed. The order of appearance and the content of the softkey
menus depend on the measurement channel 1 and measurement channel 2 configu-
ration.
KEYSTROKES
SCPI COMMAND
CAL,
(Fault Location)
Default Cal
SENS[1|2]:CORR:CSET DEF; *WAI
Full Band Cal
SENS[1|2]:CORR:EXT [ON|OFF]
Measure Standard
1
SENS[1|2]:CORR:COLL STAN[1|2|3];*WAI
Velocity Factor
SENS[1|2]:CORR:RVEL:COAX <value>
Cable Loss
SENS[1|2]:CORR:LOSS:COAX <value>
Calibrate Cable
(menu selection only)
Specify Length
SENS[1|2]:CORR:LENG:COAX <value>; *WAI
Measure Cable
SENS[1|2]:CORR:RVEL; *WAI
Multi Peak
(menu selection only)
Multi Peak Corr on OFF
SENS[1|2]:CORR:PEAK:COAX ON
Multi Peak Threshold
SENS[1|2]:CORR:THRESHOLD:COAX -10.00
Connector Values
(menu selection only)
Connector Length
SENS[1|2]:CORR:LENG:CONN <value>
Connector C
SENS1:CORR:CAP:CONN <value>
More Cal
(see
)
(menu selection only)
Summary of Contents for 8712ES
Page 11: ...1 1 1 Introduction to GPIB Programming ...
Page 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Page 36: ...3 1 3 Passing Control ...
Page 39: ...4 1 4 Data Types and Encoding ...
Page 46: ...5 1 5 Using Status Registers ...
Page 71: ...6 1 6 Trace Data Transfers ...
Page 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Page 99: ...7 1 7 Using Graphics ...
Page 105: ...8 1 8 Front Panel Keycodes ...
Page 111: ...9 1 9 Introduction to SCPI ...
Page 129: ...10 1 10 Menu Map with SCPI Commands ...
Page 268: ...12 1 12 SCPI Conformance Information ...
Page 290: ...13 1 13 SCPI Error Messages ...