![Agilent Technologies 8712ES Programmer'S Manual Download Page 213](http://html.mh-extra.com/html/agilent-technologies/8712es/8712es_programmers-manual_2867956213.webp)
Programmer’s Guide
11-15
SCPI Command Summary
SCPI Device Command Summary
Table 11-6
DIAGnostic (2 of 4)
SUBSYSTEM COMMANDS
FORM
DESCRIPTION
DIAGnostic:COMMunicate:LAN:SEND
<IP_address>,<port_num>,<string>
,
<
timeout>
NR1,
STRING
Instructs the analyzer to open a
socket to the specified IP
address and port number, and
send the string specified.
<timeout>
is an integer
<num>
in the range 0-75 specifying the
number of seconds allocated for
a successful transmission. If 0 is
specified, a minimum interval of
0.10 seconds is used. If
<timeout> is not specified, the
default interval of 75 seconds is
used.
DIAGnostic:COMMunicate:LA
N:SEND
? returns
0
if the last
socket connection was
successful, and
-1
if the last
socket connection failed. This
may not be the last socket
connection made by the IBASIC
program. Multiple socket
connections are possible, and
telnet sessions may be
interspersed among
programmed socket connections.
DIAGnostic:MDISplay[1|2]
:CORRection
<I_Direct|I_SRcmatch|I_TRacking
|I_Response|I_LDmatch|I_ISolation
|M_Direct|M_ISolation|M_SRcmatch
|M_TRacking|M_Response|M_LDmatch
|M_XScalar|C_Direct|C_SRcmatch
|C_RTracking|C_LDmatch
|C_TTracking|C_Isolation>
command
only
Return to measurement mode
and auto scales after viewing
calibration uncertainties.
Summary of Contents for 8712ES
Page 11: ...1 1 1 Introduction to GPIB Programming ...
Page 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Page 36: ...3 1 3 Passing Control ...
Page 39: ...4 1 4 Data Types and Encoding ...
Page 46: ...5 1 5 Using Status Registers ...
Page 71: ...6 1 6 Trace Data Transfers ...
Page 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Page 99: ...7 1 7 Using Graphics ...
Page 105: ...8 1 8 Front Panel Keycodes ...
Page 111: ...9 1 9 Introduction to SCPI ...
Page 129: ...10 1 10 Menu Map with SCPI Commands ...
Page 268: ...12 1 12 SCPI Conformance Information ...
Page 290: ...13 1 13 SCPI Error Messages ...