![Agilent Technologies 8712ES Programmer'S Manual Download Page 79](http://html.mh-extra.com/html/agilent-technologies/8712es/8712es_programmers-manual_2867956079.webp)
Programmer’s Guide
6-9
Trace Data Transfers
Using Binary Data Encoding
The analyzer stores its internal data with approximately 5 significant
digits of resolution. Using REAL,32 or ASCII,5 format provides sufficient
precision for data transfers. However, REAL,64 may be necessary when
using a programming language which does not support IEEE 32-bit
floating point.