Programmer’s Guide
10-15
Menu Map with SCPI Commands
Menu Map for 8712ET/ES and 8714ET/ES
Table 10-3
BEGIN
Functions, 8712ES and 8714ES (3 of 3)
1. The active measurement channel configuration determines the order of appearance
and the content of the softkey menus.
2. Option 100 (SRL and Fault Location) only.
KEYSTROKES
SCPI COMMAND
BEGIN, Cable,
1,2
(continued)
SRL
, (continued)
Connector Model
(menu selection only)
Measure Connector
SENS[1|2]:CORR:MODEL:CONN
Connector Length
SENS[1|2]:CORR:LENG:CONN <num>
Connector C
SENS[1|2]:CORR:CAP:CONN <num>
Z Cutoff Frequency
SENS:FREQ:ZST <num>
Auto Z ON off
SENS1:FUNC:SRL:MODE [MANUAL|AUTO]
Manual Z
SENS1:FUNC:SRL:IMP <num>
Connector Fault
SENS1:FUNC:FAULT:CONN; *WAI
SRL Cable Scan
SENS[1|2]:FUNC:SRL:SCAN; *WAI
Number of Points
SENS[1|2]:SWE:POIN <num>;*WAI
Autost
(menu selection only)
User Begin on OFF
(menu selection only)
Summary of Contents for 8712ES
Page 11: ...1 1 1 Introduction to GPIB Programming ...
Page 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Page 36: ...3 1 3 Passing Control ...
Page 39: ...4 1 4 Data Types and Encoding ...
Page 46: ...5 1 5 Using Status Registers ...
Page 71: ...6 1 6 Trace Data Transfers ...
Page 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Page 99: ...7 1 7 Using Graphics ...
Page 105: ...8 1 8 Front Panel Keycodes ...
Page 111: ...9 1 9 Introduction to SCPI ...
Page 129: ...10 1 10 Menu Map with SCPI Commands ...
Page 268: ...12 1 12 SCPI Conformance Information ...
Page 290: ...13 1 13 SCPI Error Messages ...