![Agilent Technologies 8712ES Programmer'S Manual Download Page 307](http://html.mh-extra.com/html/agilent-technologies/8712es/8712es_programmers-manual_2867956307.webp)
13-18
Programmer’s Guide
SCPI Error Messages
Query Errors
Table 13-4
SCPI Query Errors
Error
Number
Error Description
−
400
Query error — This is the generic query error for devices that cannot detect
more specific errors. This code indicates only that a Query Error has
occurred.
−
410
Query
INTERRUPTED
— Indicates that a condition causing an
INTERRUPTED
Query error occurred; for example, a query followed by
DAB
or
GET
before a response was completely sent.
−
420
Query
UNTERMINATED
— Indicates that a condition causing an
UNTERMINATED
Query error occurred; for example, the device was
addressed to talk and an incomplete program message was received.
−
430
Query
DEADLOCKED
— Indicates that a condition causing a
DEADLOCKED
Query error occurred; for example, both input buffer and output buffer are
full and the device cannot continue.
−
440
Query
UNTERMINATED
after indefinite response — Indicates that a query
was received in the same program message after an query requesting an
indefinite response was executed.
Summary of Contents for 8712ES
Page 11: ...1 1 1 Introduction to GPIB Programming ...
Page 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Page 36: ...3 1 3 Passing Control ...
Page 39: ...4 1 4 Data Types and Encoding ...
Page 46: ...5 1 5 Using Status Registers ...
Page 71: ...6 1 6 Trace Data Transfers ...
Page 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Page 99: ...7 1 7 Using Graphics ...
Page 105: ...8 1 8 Front Panel Keycodes ...
Page 111: ...9 1 9 Introduction to SCPI ...
Page 129: ...10 1 10 Menu Map with SCPI Commands ...
Page 268: ...12 1 12 SCPI Conformance Information ...
Page 290: ...13 1 13 SCPI Error Messages ...