Programmer’s Guide
10-25
Menu Map with SCPI Commands
Menu Map for 8712ET/ES and 8714ET/ES
Table 10-10
CAL
,
Cal Check
Functions, 8712ET/ES and 8714ET/ES
1. Only those calibration checks valid for the current measurement type are active,
other choices are dimmed. For example,
Transmission Tracking
is not valid for reflec-
tion measurements.
KEYSTROKES
SCPI COMMAND
CAL
,
Cal Check
(menu selection only)
Do Cal Check
SENS1:CORR:COLL:IST OFF;METH VERIFY;*WA
I
Measure Standard
SENS[1|2]:CORR:COLL:VER:REFL
STAN[1|2|3];*WAI;
View Cal Check
1
(menu selection only)
Directivity
DIAG:MDIS[1|2]:CORR C_DIRECT; *WAI
Source Match
DIAG:MDIS[1|2]:CORR C_SRCMATCH; *WAI
Reflection Tracking
DIAG:MDIS[1|2]:CORR C_RTRACKING; *WAI
Load Match
DIAG:MDIS[1|2]:CORR C_LDMATCH; *WAI
Transmissn
Tracking
DIAG:MDIS[1|2]:CORR C_TTRACKING; *WAI
Isolation
DIAG:MDIS[1|2]:CORR C_ISOLATION; *WAI
Restore Meas
DIAG:MDIS[1|2]:REST; *WAI
Summary of Contents for 8712ES
Page 11: ...1 1 1 Introduction to GPIB Programming ...
Page 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Page 36: ...3 1 3 Passing Control ...
Page 39: ...4 1 4 Data Types and Encoding ...
Page 46: ...5 1 5 Using Status Registers ...
Page 71: ...6 1 6 Trace Data Transfers ...
Page 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Page 99: ...7 1 7 Using Graphics ...
Page 105: ...8 1 8 Front Panel Keycodes ...
Page 111: ...9 1 9 Introduction to SCPI ...
Page 129: ...10 1 10 Menu Map with SCPI Commands ...
Page 268: ...12 1 12 SCPI Conformance Information ...
Page 290: ...13 1 13 SCPI Error Messages ...