5-22
Status Listings
OSA Warnings
5058
desc = “Out of memory”
help = “The OSA has run out of execution memory. An internal function was
aborted due to lack of execution memory. The correction for this problem is to cycle
power.”
5059
desc = “Trajectory align: marginal input signal”
help = “The trajectory align procedure was aborted due to an input signal with
marginal amplitude. The input signal was large enough when the function began
it’s operation but later became too small. This usually happens when the signal is
on the edge of being too small. To correct this problem increase the input signal
level. If this is not possible then a slight increase in signal level can sometimes be
achieved by cleaning optical fiber connectors.”
5060
desc = “Trajectory align failed”
help = “The trajectory align procedure failed. The trajectory adjustments computed
were invalid. This is usually occurs after the OSA has received a large mechanical
shock. To correct this problem try the trajectory align procedure again. If the error
persists, contact the Agilent support center nearest your location.”
5061
desc = “Invalid marker trace”
help = “The marker could not be positioned. The marker is on a trace which does
not contain any valid data. This is sometimes caused by a trace math result which is
invalid.”
5062
desc = “ADC Triggered Sweep Too Fast”
help = “A sweep was taken with one of the ADC trigger modes enabled. The
sweep rate was too fast to allow trace data to be acquired for every wavelength.
To resolve this problem increase sweep time or increase the frequency the external
trigger input signal.”
5063
desc = “Key disabled during applications”
help = “The key you pressed is not active while an application is running. Exiting
the application should re-enable the key.”
5064
desc = “Invalid measurement file”
help = “An attempt was made to recall a measurement file which is contains invalid
or corrupt data.”
5065
desc = “Error detected in DSP sub-system”
help = “The Digital Signal Processor has reported an unexpected error. Please
record the hexadecimal number listed with the error and cycle power. If the error
persists, contact the Agilent support center nearest your location.”
Table 5-6. OSA Warnings (6 of 12)
Error Number
Error Description [description/explanation/examples]
Summary of Contents for 8614 B Series
Page 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Page 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Page 13: ...1 5 Getting Started Product Overview ...
Page 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 26: ...1 18 Getting Started The Softkey Panels ...
Page 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Page 40: ...1 32 Getting Started Product Options and Accessories ...
Page 75: ...3 Function Reference ...
Page 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Page 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Page 271: ......