3-95
Function Reference
Wavelength Offset
Key Path
Markers > More Marker Functions > Line Marker Menu > Wavelength Line Mkr 1| 2
Remote Commands
(Wavelength Line Marker 1):CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]
:SRANge:LOWer
(Wavelength Line Marker 2):CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]
:SRANge:UPPer
(Sweep Limit):SENSe:WAVelength:SRANge:UPPer
Wavelength Offset
Specifies the wavelength offset. This is an offset between the measured wavelength and the dis-
played wavelength. You can calibrate the TLS with a wave meter and correct for any offset.
Entering a value in the wavelength offset has no affect on the trace.
Key Path
Wavelength > Wavelength Setup > Wavelength Offset
Remote Commands
SENSe:WAVelength:OFFSet
Wavelength Setup
Accesses the Wavelength Setup panel:
• center wavelength step size
• user wavelength calibration date
• wavelength calibration
• wavelength offset
• wavelengths referenced in
• wavelength units
Key Path
Wavelength > Wavelength Setup
Remote Commands
CALibrate:WAVelength:STATe ON
CALibrate:WAVelength:STATe OFF
SENSe:WAVelength:OFFSet
SENSe:WAVelength:CENTer:STEP:INCRement
Summary of Contents for 8614 B Series
Page 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Page 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Page 13: ...1 5 Getting Started Product Overview ...
Page 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 26: ...1 18 Getting Started The Softkey Panels ...
Page 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Page 40: ...1 32 Getting Started Product Options and Accessories ...
Page 75: ...3 Function Reference ...
Page 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Page 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Page 271: ......