3-83
Function Reference
Switch Path No Auto Align
N o t e
The auto align now will overwrite any previous align data. To preserve current auto align data,
select Switch Path No Auto Align.
The data returned by the alignment is stored in both the 9 µm and the 50 µm table. With the data
stored on both tables, the centering for the 50 µm path is improved due to the increased resolu-
tion bandwidth of the 9 µm path. Once the align is complete or if you select No Auto Align, the
instrument will be ready to detect data via the 50 µm path.
Key Path
Appl’s > Measurement Modes > Filter Mode > Switch Path Auto Align Now
Related Functions
Switch Path No Auto Align
Remote Commands
ROUTe:PATH:EXTernal
CALibration:ALIGn:FILTer
Switch Path No Auto Align
For 86146B Filter Mode only.
Selects the 9 µm filter mode path. You should select this function if the instrument has not be
moved, subjected to temperature changes >2° C, turned off and warmed up for at least an hour
at the start of each day or to preserve previous align data.
Key Path
Appl’s > Measurement Modes > Filter Mode > Switch Path Auto Align Now
Related Functions
Switch Path Auto Align Now
Remote Commands
ROUTe:PATH:EXTernal
Summary of Contents for 8614 B Series
Page 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Page 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Page 13: ...1 5 Getting Started Product Overview ...
Page 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 26: ...1 18 Getting Started The Softkey Panels ...
Page 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Page 40: ...1 32 Getting Started Product Options and Accessories ...
Page 75: ...3 Function Reference ...
Page 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Page 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Page 271: ......