Agilent 86038B Photonic Dispersion and Loss Analyzer,
Second Edition
265
Property Value
InstrumentObjects.ODASensitivity.eIFBandwidth
VB.NET Syntax
pdlaClient.Sensitivity.IFBandwidth = _
InstrumentObjects.ODASensitivity.eIFBandwidth.IFBW_100Hz
VB 6.0 Syntax
pdlaClient.Sensitivity.IFBandwidth=eIFBandwidth_IFBW_100H
z
SamplesPerPoint Property
Sets the number of amplitude and phase measurements
(samples) that are taken at each measurement wavelength
(trace point).
In Swept measurement mode, the recovered modulation
amplitude and phase are sampled periodically as the
wavelength is swept. As a result, the wavelength resolution of
the measurement is limited by wavelength range/(number of
points per sweep – 1). Note that wavelength resolution is also
affected by the modulation frequency.
In Stepped measurement mode, you select how many samples to
take at each measurement wavelength. The multiple samples are
averaged to generate a single trace point. A larger number of
points means more averaging, suitable for reducing trace noise
in cases of high insertion loss test devices.
Property Value
Integer - The number of samples to take at each point.
Related Command
Sensitivity.ValidSamplesPerPoint method
VB.NET Syntax
pdlaClient.Sensitivity.SamplesPerPoint = 201
VB 6.0 Syntax
pdlaClient.Sensitivity.SamplesPerPoint = 201
Summary of Contents for 86038B
Page 1: ...Agilent 86038B Photonic Dispersion and Loss Analyzer User s Guide ...
Page 4: ...4 ...
Page 20: ...20 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 34: ...34 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition Figure 2 b Rear Panel ...
Page 78: ...78 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 92: ...92 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 202: ...202 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition End Sub ...
Page 348: ...348 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 349: ......