MeasurementRange Commands
The MeasurementRange property of the remote client object
returns a measurement range object that contains the
functionality of the front panel Measurement Range panel.
ValidStartStop Method
Returns valid minimum and maximum wavelength values for
the measurement.
Parameters
ByRef Minimum As Double - minimum wavelength accepted, in
nm.
ByRef Maximum As Double - maximum wavelength accepted, in
nm.
Related Command
MeasurementRange.XStart property, MeasurementRange.XStop
property
VB.NET Syntax
Dim minWL As Double
Dim maxWL As Double
pdlaClient.MeasurementRange.ValidStartStop (minWL, maxWL)
VB 6.0 Syntax
Dim minWL As Double
Dim maxWL As Double
Call pdlaClient.MeasurementRange.ValidStartStop (minWL,
maxWL)
XStart Property
Sets and reads the start wavelength for the next measurement.
Property Value
Double - The start wavelength in nanometers.
Summary of Contents for 86038B
Page 1: ...Agilent 86038B Photonic Dispersion and Loss Analyzer User s Guide ...
Page 4: ...4 ...
Page 20: ...20 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 34: ...34 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition Figure 2 b Rear Panel ...
Page 78: ...78 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 92: ...92 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 202: ...202 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition End Sub ...
Page 348: ...348 Agilent 86038B Photonic Dispersion and Loss Analyzer Second Edition ...
Page 349: ......