Installation
8-27
7.
Strip off approximately 6.3 mm (1/4 in.) of insulation from each of the
two wire ends to allow the bared end to be soldered onto the connector
contacts.
8.
Solder the two wire ends and the drain wire to the respective terminal
contacts as indicated below.
9.
Connect:
PINK WIRE (positive) ---------->pin A
DRAIN WIRE (GND) ---------- >pins B, C, and D
BLUE WIRE (negative) ---------- >pin E
10. Slide the connector locking ring forward over the connector as far as
possible.
11. Screw the backshell onto the connector and tighten.
12. Gently, slide the cable clamp over the folded back braided shield and
screw the clamp onto the backshell. Do not twist the connector; twist
the cable clamp.
13. Tighten down both cable clamps uniformly. Trim off excess braided
shield.
14. After you create the cable, screw the connector to the M3199AI option
J03 (78581-60350).
15. Plug the RJ45 end into the appropriate port on the ACC. Refer to the
following figures.