Installation
8-7
Environmental
Conditions
The ACC mounting location must have adequate ventilation and air
circulation, and be in a relatively dust/lint-free, static-free environment.
The operating environmental characteristics of the ACC mounting location
must conform to the environmental specifications listed.
Temperature Range (ambient): Operating 0
o
C to 55
o
C (32
o
F to 131
o
F)
Humidity: Up to 95% RH
Altitude: Operating to 4600 meters (15,000 ft.)
Important Instructions
•
Allow a minimum of 5.08 cm (2 in.) of clearance on top and bottom.
•
Allow a minimum of 1 m (3 ft.) of free space in for service access.
•
Do not mount in an unventilated area such as a closet or console.
•
Mount within 1.5 m (5 ft.) of an A.C. power source.
•
When disconnecting from power, pull the cord cap.
•
Power Requirements:
100 - 240V at 50 - 60 Hz
•
Power Consumption:
80 VA.
•
Heat Dissipation:
51.6 kg/cal/hr (204.8 BTU/hr.).
•
Line Conditioning Requirements: None
•
Weight:
14.1 lb. (6.4 kg).