Application Tutorials
Tutorial 3: Measure Extinction Ratios on Low-Level Signals
Measure the Extinction Ratio
8
To measure the extinction ratio, press:
Measure
,
NUMBER SAMPLES
and enter
20000
,
EXTINCT RATIO
Refer to the following figure.
Extinction Ratio Measurement on a Low-Level Signal with Eye Filtering On
Display the Signal in Optical Units
9
To display the signal in optical units, press:
Trg,Cal
,
CH1 EXT ATTEN
and enter
30
V/Watt