3-45
Eye-Diagram Analyzer Reference
Saving to Mass Storage
To Erase a Custom
Prefix
1
To erase a custom prefix, press:
MENU
,
page 1 of 2
,
States
,
more 1 of 2,
mass storage
,
more
,
CHANGE
PREFIX
Repeatedly press the
⇓
step key until the cursor is located at the start of the
custom text.
2
Clear the text by pressing:
CLEAR TO END
,
ENTER LINE
To Save an
Instrument State
1
Set the eye-diagram analyzer to the desired settings.
2
Insert a ROM card into the front-panel card slot.
3
Save the instrument state by pressing:
Mass Storage
,
SAVE STATE
To Load a State or
Mask/Limit Line
1
Insert a ROM card into the front-panel card slot and then press:
Mass Storage
Rotate the front-panel knob to highlight the desired file.
The prefix "S_" identifies instrument state files.
The prefix "M_" identifies mask/limit line files.
2
Load the file by pressing:
LOAD FILE
Summary of Contents for 71501D
Page 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Page 7: ...1 Getting Started...
Page 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Page 35: ...2 Application Tutorials...
Page 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Page 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Page 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Page 81: ...3 Eye Diagram Analyzer Reference...
Page 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Page 151: ...4 Programming Commands...
Page 163: ...5 Specifications and Characteristics...
Page 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...