1-19
Getting Started
Steps for Setting Up Eye-Diagram Analysis
Step 3. Complete the Installation Using the
Screen Instructions
• If the displayed screen looks like the figure on the left side of this page, your
system has been previously configured. Press CONT and then continue with
"Step 4. Connect the front-panel cables". However, if you wish to reconfigure
your system, press RECNFIG, and continue with the following explanation of
the self-guided screens.
• If the screen looks like the figure on the right side of this page, the first self-
guided screen is displayed. Perform the following steps:
1
Press 7084X if you are using an 70841A/B pattern generator. If using an
70843A, or any other pattern generator, press MANUAL.
2
If you press 7084X, the screen shown on the following page is displayed. The
program automatically determines and displays the pattern generator module’s
HP-MSIB address. For most installations, press CONT. To manually enter the
HP-MSIB address, use the displayed softkeys.
Example of a Configuration Previously Done
Summary of Contents for 71501D
Page 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Page 7: ...1 Getting Started...
Page 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Page 35: ...2 Application Tutorials...
Page 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Page 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Page 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Page 81: ...3 Eye Diagram Analyzer Reference...
Page 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Page 151: ...4 Programming Commands...
Page 163: ...5 Specifications and Characteristics...
Page 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...