4263B
5.
Set
the
test
signal
frequency
to
100
Hz
using
.
6.
Record
the
Multimeter
reading
on
the
calculation
sheet.
7.
Calculate
the
test
result
according
to
the
calculation
sheet,
and
record
the
result
into
the
performance
test
record.
8.
Change
the
test
signal
settings
by
using
and
to
perform
this
test
for
all
settings
listed
in
T
able
9-3 .
T
able
9-3.
Level
A
ccuracy
T
est
Settings
T
est
Signal
Level
T
est
Signal
Frequency
50
mV
100
Hz
20
kHz
1
100
kHz
250
mV
100
Hz
20
kHz
1
100
kHz
1
V
100
Hz
20
kHz
1
100
kHz
1
Option
002
only
.
9-8
Maintenance
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 4263B
Page 14: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 15: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 17: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 18: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 19: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 21: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 199: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 209: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 241: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 295: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 307: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 317: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...