Troubleshooting
5
16900A, 16902A, and 16903A Service Guide
35
Figure 8
16903A Display Troubleshooting Flowchart
3
Is there
any boot dialog at
any time?
Replace the PCI bridge
board.
Connect a known good external
monitor. NOTE: press and hold
"Touch off" to switch between the
internal and external monitors.
Proceed to boot up
troubleshooting.
4
Yes
Login to Windows and check display
settings. Set the default to display 1
(internal display) at 800x600 pixels
(16-bit color depth).
Check all cables to the front panel
assembly. Check all cables to the
CPU motherboard. Ensure the PCI
bridge board is fully seated. Reseat
as necessary.
No
E
Is the
display screen readable
on the external
monitor?
No
Yes
Replace the CPU
motherboard.
1
Replace the MIB.
D
Try , then .
Return to between
each.
A
B
1
B
A
Is the
built-in display
completely
dark?
No
Yes
C
Try & return to .
1
Replace the LCD
panel.
C
D
E
Try , and then .
Return to between each.
1
Summary of Contents for 16900A
Page 1: ...Agilent Technologies Agilent 16900A 16902A and 16903A Logic Analysis System Service Guide...
Page 6: ...6 16900A 16902A and 16903A Service Guide...
Page 14: ...14 16900A 16902A and 16903A Service Guide 1 General Information...
Page 18: ...18 16900A 16902A and 16903A Service Guide 2 Preparing for Use...
Page 22: ...22 16900A 16902A and 16903A Service Guide 3 Testing Performance...
Page 86: ...86 16900A 16902A and 16903A Service Guide 6 Replacing Assemblies...
Page 102: ...102 16900A 16902A and 16903A Service Guide 7 Replaceable Parts...