82
16900A, 16902A, and 16903A Service Guide
6
Replacing Assemblies
To remove and replace the line filter assembly
Perform previous procedures:
•
“To prepare the instrument for disassembly"
•
“To remove and replace the cover"
1
Disconnect the power cable from the power supply.
2
Using a Torx T10 screwdriver, remove the 2 screws
securing the power socket to the frame and the 1 screw
to the ground wire.
3
Reverse this procedure to replace.
Line Filter Assembly
(16900A & 16902A)
Ground Wire
T10 Screw
T10 Screws (2)
Line Filter Assembly
(16903A)
Ground Wire
T10 Screw
T10 Screws (2)