![Agilent Technologies 16900A Service Manual Download Page 58](http://html1.mh-extra.com/html/agilent-technologies/16900a/16900a_service-manual_2867161058.webp)
58
16900A, 16902A, and 16903A Service Guide
6
Replacing Assemblies
To remove and replace optional interface cards
1
Perform previous procedures:
•
“To prepare the instrument for disassembly"
•
“To remove and replace the cover"
•
“To remove and replace the CPU tray"
2
Using a T10 screwdriver, remove the screw securing the
interface card to the rear panel of the CPU tray.
3
Swing the plate over the tall slot(s) out of the way as
shown in the following figure.
4
Lift the card out of the socket to remove.
5
To install a card, ensure that the tab at the bottom of the
card is aligned with the slot in the bottom of the CPU
tray between the motherboard and the rear panel.
6
Press the card securely into the socket on the
motherboard.
7
For cards in the tall slot(s), swing the plate back into
place.
8
Replace the screw(s).