16900A, 16902A, and 16903A Service Guide
6
Replacing Assemblies
To remove and replace the cover
1
Perform previous procedures:
•
“To prepare the instrument for disassembly"
2
Using a Torx T15 screwdriver, remove the handle
assembly.
3
Using a Torx T10 screwdriver, remove 3 screws that
secure the cover to the chassis.
4
Slide the cover back and up to remove.
5
Reverse this procedure to replace the cover.
T-15 Screws
T10 Screws
C A U T I O N
When reinstalling the handle assembly, ensure that the screws are
torqued to 2.372 Newton meters (21 inch pounds) so that they do
not work themselves loose.