![Agilent Technologies 16900A Service Manual Download Page 73](http://html1.mh-extra.com/html/agilent-technologies/16900a/16900a_service-manual_2867161073.webp)
Replacing Assemblies
6
16900A, 16902A, and 16903A Service Guide
73
5
When re- assembling, feed the keypad cable and the power
On|Off cable through the LCD bracket.
6
Align the 6 long pins on the LCD bracket with the guide
pins sticking through the keypad board.
7
Press on the LCD bracket assembly firmly to ensure the
LCD display is attached to the touch screen.
8
Using a Torx T10 screwdriver, screw the 6 screws into the
LCD bracket to secure it to the front frame.
T10 Screws (6)
Slots
LCD Bracket
Pink LCD
Cable
Liquid Crystal Display
(LCD)
LCD Cable
Touch
Screen
Touch
Screen
Cable
Long Alignment Pins (6)