Chapter 1: Agilent Technologies 16712A 128K Sample Logic Analyzer
Importing Netlist and ASCII Files
“Performing Clock Setup (State only)” on page 42
Setting the Pod Threshold
The pod threshold is a voltage level which the data must cross before
the analyzer recognizes it as a change in logic levels. You can specify a
threshold level for each pod. The level specified for the master pod that
includes the clock is also used for the clock threshold.
Standard
1. In the
Format
tab, select the threshold button located just below the pod
name.
2. In the Pod threshold dialog, choose one of the standard threshold options:
•
TTL -- The threshold level is +1.50 volts.
•
LVTTL -- The threshold level is +1.40 volts.
•
SSTL2 -- The threshold level is +1.50 volts.
•
SSTL3 -- The threshold level is +1.25 volts.
•
LVCMOS 1.5v -- The threshold level is +0.75 volts.
•
LVCMOS 1.8v -- The threshold level is +0.90 volts.
•
LVCMOS 2.5v -- The threshold level is +1.25 volts.
•
LVCMOS 3.3v -- The threshold level is +1.65 volts.
•
CMOS 5.0v -- The threshold level is +2.50 volts.
•
ECL -- The threshold level is -1.3 volts.
•
LVPECL -- The threshold level is 2.00 volts.
3. If you don't want the change to apply to all pods, deselect the checked box
next to
Apply settings to all pods
.
4. Select the
Close
button.