![Agilent Technologies 16712A Help Manual Download Page 70](http://html1.mh-extra.com/html/agilent-technologies/16712a/16712a_help-manual_2867831070.webp)
69
Chapter 1: Agilent Technologies 16712A 128K Sample Logic Analyzer
Importing Netlist and ASCII Files
to
Master/Slave
or
Demultiplex
. The Pod Clock field is located just
below the Pod Threshold in
Format
.
The Pod Clock field is set to
Master Clk
by default. Use the Pod Clock
field to indicate if the pod's data is to be strobed into memory by the
master clock, slave clock, or both, in the demultiplex mode.
When the Pod Clock is set to
Demultiplex
, only one pod of a
pod pair
is usable. That pod latches data on both the master and slave clocks, so
it appears twice in the label area. To select which pod of a pod pair you
want to demultiplex, select the Pod Field.
See Also
“Performing Clock Setup (State only)” on page 42
“Clock Modes (State only)” on page 43 for details on slave and demultiplex
clocks
“Pod Selection” on page 69
Pod Selection
The Pod field in
Format
identifies which pod of a
pod pair
the settings
of the bit assignment field, pod threshold field, and pod clock fields
effect. Most of the time it is simply informational. The exceptions are
noted below.
Half-Channel Mode
In the half-channel mode, the Pod field becomes a button that you can
use to select which pod of a pod pair all pod settings apply to. In the
full-channel modes, this field is simply an identifier and is not
selectable.
Demultiplex Clock Mode
In the demultiplex clock mode, use the pod field to select which pod of
a pod pair is to be used to sample data. In the master or slave clock
modes, this field is simply an identifier and is not selectable.
See Also
“Setting the Acquisition Mode” on page 42
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com