Chapter 1: Agilent Technologies 16712A 128K Sample Logic Analyzer
Importing Netlist and ASCII Files
channel
. An active channel looks like
.
A dash at the top of the activity indicator display indicates that the
signal connected to that channel is electrically high (above the
threshold voltage). A dash at the bottom indicates that the signal is
electrically low. An arrow indicates that the signal is transitioning.
Activity indicators are not affected by label polarity. (see page 66)
You can use these indicators to check whether there is proper probe
connection:
bits
that are stuck high or low may not be properly
connected. You can also verify that the signals in your target system are
active: bits that are stuck high or low are not crossing the threshold
voltage.
Activity indicators are not displayed during an
acquisition
.
See Also
“Setting the Pod Threshold” on page 70
Logic Analysis System and Measurement Modules Installation Guide
for details on probing
Assigning Pods to the Analyzers
The
Pod Assignment...
button under
Format
can be used to start the
second analyzer on the
module
and to assign pod pairs.
To Assign Pods to an Analyzer
1. In the
Format
tab, select the
Pod Assignment...
button.
The Pod Assignment window appears.
2. Drag a pod pair and drop it below the analyzer that you want to assign it to.
3. For pods that you do not want to use,
drag and drop
them in the
Unassigned Pods area. This preserves memory depth when
count
is turned
on.
NOTE:
When both analyzers are turned on, pods 1/2 and 3/4 of the master card can
not be assigned to the same analyzer.