Detector
Variable Wavelength Detector (VWD)
Features (G1314F)
The Agilent variable wavelength detectors described in this manual is
designed for highest optical performance, GLP compliance and easy
maintenance with:
•
data rate up to 80 Hz for standard-HPLC
•
deuterium lamp for highest intensity and lowest detection limit over a
wavelength range of 190 to 600 nm,
•
optional flow-cell cartridges (standard 10 mm, 14 µL; high pressure 10 mm,
14 µL; micro 3 mm, 2 µL; semi-micro 6 mm, 5 µL) are available and can be
used depending on the application needs (other types may be introduced
later),
•
easy front access to lamp and flow cell for fast replacement,
•
electronic identification of flow cell and lamp with RFID (Radio Frequency
Identification) tag for unambiguous identification,
•
lamp information: part number, serial number, production date,
ignitions, burn time
•
cell information: part number, serial number, production date, nominal
path length, volume, maximum pressure
•
built-in electronic temperature control (ETC) for improved baseline
stability, and
•
built-in holmium oxide filter for fast wavelength accuracy verification.
“Performance Specifications (G1314F)”
N O T E
This detectors cannot be operated with a G1323B Control Module. Use the Instant Pilot
(G4208A) as local controller.