Agilent 1260 Infinity Quaternary LC VL Manual and Quick Reference
61
Optimization of the Agilent 1260 Infinity Quaternary LC VL
3
Optimizing the Thermostatted Column Compartment
Optimizing the Thermostatted Column Compartment
For best performance results of the column compartment:
•
Use short connection capillaries and place them close to the heat
exchanger. This will reduce heat dissipation and external band-broadening.
•
Use the left heat exchanger for small volume columns, for example, 2 –
3 mm i.d. columns at flow rates of less than 200 µL/min.
•
For even lower band-broadening, the heat exchanger can be by-passed and
the column is placed well between the heat exchanger fins.
•
Keep the left and right heat exchanger temperature the same unless you do
specific applications.
•
Assure that the front cover is always closed.