Agilent 1260 Infinity Quaternary LC VL Manual and Quick Reference
87
System Setup and Installation
4
Installing the Modules
1
Open the purge valve of the pump
2
Set the flow rate to 5 mL/min.
3
Select channel A1
4
Turn the flow on
5
Observe if the solvent in the tubing of channel A1 is advancing towards the
pump. If it isn’t, disconnect the solvent tubing from the solvent selection
valve, attach a syringe with a syringe adapter and pull the liquid through
the degasser. Reattach the tubing to the solvent selection valve.
6
Pump 30 mL isopropanol to remove residual air bubbles.
7
Switch to the next solvent channel and repeat steps 5 and 6 until all
channels have been purged.
8
Turn the flow off and close the purge valve.