Introduction
Detector
Max-Light Cartridge Flow Cell
The detector allows easy access to flow cells via a cartridge. A variety of
optional flow cells can be inserted using the same quick, simple mounting
system. A Max-Light Cartridge Cell (10 mm, V(
σ
) 1.0 µL) and a Max-Light
Cartridge Cell (60 mm, V(
σ
) 4 µL) are available. For testing of the detector, a
Max-Light Cartridge Test Cell is available.
The optical principle of the Max-Light Cartridge cell is based on opto-fluidic
waveguides. Nearly 100 % light transmission is achieved by utilizing total
internal reflection in a non-coated silica fiber. Compromising refractive index
and thermal effects are almost completely eliminated, resulting in significantly
less baseline drift.
Figure 13
Max-Light Cartridge Flow Cell
Slit
The fixed slit combines the required optical functions - slit and shutter - in a
simple and compact component. The slit width is directly controlled by the
micro-processor of the instrument and is fixed to 4 nm.