Agilent 1260 Infinity Quaternary LC VL Manual and Quick Reference
41
Specifications
2
Performance Specifications
Performance Specifications (G1316A)
Table 5
Performance Specifications Thermostatted Column Compartment
Type
Specification
Comments
Temperature range
10 degrees below ambient to 80 °C
up to 80 °C: flow rates up to 5 mL/min
Temperature stability
± 0.15 °C
Temperature accuracy
± 0.8 °C
± 0.5 °C
With calibration
Column capacity
Three 30 cm
Warm-up/cool-down
time
5 minutes from ambient to 40 °C
10 minutes from 40 – 20 °C
Dead volume
3 µL left heat exchanger
6 µL right heat exchanger
Communications
Controller-area network (CAN), RS-232C, APG Remote: ready, start,
stop and shut-down signals, LAN via other 1260 Infinity module
Safety and maintenance
Extensive diagnostics, error detection and display (through Instant
Pilot and Agilent data system), leak detection, safe leak handling,
leak output signal for shutdown of pumping system. Low voltages
in major maintenance areas.
GLP features
Column-identification module for GLP documentation of column
type.
Housing
All materials recyclable