80
Chapter 3
Operation
Using an E4407B Spectrum Analyzer or 7405A EMC Analyzer
O
p
er
ati
o
n
Using an E4407B Spectrum Analyzer
or 7405A EMC Analyzer
Preliminary Operation
Compatibility Reference
Refer to
on page 27 for information relating to retrofit
requirements for E4407B or E7405A analyzer. The E4407B is used in the following
procedure.
Equipment Connection
Connect the equipment as shown in
and use the following checklist to verify
the connections.
• 11974 LO IN to E4407B LO OUTPUT (SMA cable, 5061-5458)
• 11974 IF OUT to E4407B IF INPUT (SMA cable, 5061-5458)
• 11974 TUNE IN to E4407B Preselector Tune Output (BNC cable, 8120-1839; BNC f–f
adapter, 1250-0080; BNC cable, 8120-2582)
• 11974 POWER SUPPLY to 11974-60028 PRESELECTOR POWER
Figure 3-8. 11974 Connections Using an E4407B or E7405A