20
Chapter 1
General Information
Specifications and Characteristics
G
e
ne
ral In
fo
rm
atio
n
AMPLITUDE MEASUREMENT RANGE (continued)
IF Subharmonic Response
Intercept (2nd Order)
(Response at f
IF
/2 above input signal)
f
IF
/2 = 160.7 MHz (8566B, 71000C,
PSA Series)
= 155.35 MHz (8560 Series
Portable Spectrum
Analyzer)
11974A
11974Q
11974U
11974V
+45 dBm min.
+45 dBm min.
+45 dBm min.
+30 dBm min.
Dynamic Range
Refer to
on page 25, and
INPUTS AND OUTPUTS
RF Input
Waveguide Size, Flange Type
11974A
11974Q
11974U
11974V
Maximum RF Input Level
(CW, Peak, or Average)
Match (at peak of preselector)
Size
WR-28
WR-22
WR-19
WR-15
+25 dBm
<3.0:1
Flange
UG 599/U
UG 383/U
UG 383/U-M
UG 385/U
LO Input
Connector Type
Frequency Range
11974A
11974Q
11974U
11974V
Optimal LO Input Level
Match
SMA (f)
3.3 GHz to 5.0 GHz
3.3 GHz to 5.0 GHz
4.0 GHz to 6.0 GHz
3.5 GHz to 5.3 GHz
+14.5 dBm to 16 dBm
<2.0:1 (50
Ω
)
Table 1-2 11974 Characteristics (Continued)
Summary of Contents for 11974 Series
Page 5: ...4 ...
Page 8: ...7 General Information 1 General Information ...
Page 28: ...27 Installation 2 Installation ...
Page 47: ...46 Chapter2 Installation Returning Equipment for Service Installation ...
Page 48: ...47 Operation 3 Operation ...
Page 97: ...96 Chapter3 Operation Using 8561A 62A 62B Spectrum Analyzers Operation ...
Page 98: ...97 Performance Tests 4 Performance Tests ...
Page 148: ...147 Safety and Regulatory Information 5 Safety and Regulatory Information ...