128
Chapter 4
Performance Tests
4. Displayed Average Noise Level
P
e
rf
orm
a
nc
e T
e
s
ts
14.Record the following in
Center Frequency
Displayed Average Noise Level (Marker Amplitude)
15.On the spectrum analyzer, press
SHIFT
and
SWEEP TIME
to turn off the video average
function.
16.On the spectrum analyzer, press
CENTER FREQUENCY
and enter the corresponding
value for the band listed below:
for an 11974A: 40 GHz
for an 11974Q: 50 GHz
for an 11974U: 60 GHz
for an 11974V: 75 GHz
17.Set the conversion loss to the value recorded for this frequency in
Press
SHIFT
, 5, and
MHz
, then enter the conversion loss value from
dB
to terminate the entry. Record this value in
18.On the spectrum analyzer, press
SHIFT
,
VIDEO BW
, 10, and
Hz
to activate the video
average function to average 10 traces.
19.After the video average counter reaches 10, press MARKER
NORMAL
and
PEAK SEARCH
. The marker amplitude is the displayed average noise level.
20.Record the following in
Center Frequency
Displayed Average Noise Level (Marker Amplitude)
21.On the spectrum analyzer, press
SHIFT
SWEEP TIME
to turn off the video average
function.
22.On the spectrum analyzer, press
CENTER FREQUENCY
and enter the frequency
corresponding to the maximum conversion loss measured in
23.Set the conversion loss to the value recorded for this frequency in
Press
SHIFT
5
MHz
, then enter the conversion loss value from
Press
dB
to terminate the entry. Record this value in
24.On the spectrum analyzer, press
SHIFT
,
VIDEO BW
, 10, and
Hz
to activate the video
average function to average 10 traces.
25.After the video average counter reaches 10, press MARKER
NORMAL
and
PEAK SEARCH
. The marker amplitude is the displayed average noise level.
26.Record the following in
Center Frequency
Displayed Average Noise Level (Marker Amplitude)
Summary of Contents for 11974 Series
Page 5: ...4 ...
Page 8: ...7 General Information 1 General Information ...
Page 28: ...27 Installation 2 Installation ...
Page 47: ...46 Chapter2 Installation Returning Equipment for Service Installation ...
Page 48: ...47 Operation 3 Operation ...
Page 97: ...96 Chapter3 Operation Using 8561A 62A 62B Spectrum Analyzers Operation ...
Page 98: ...97 Performance Tests 4 Performance Tests ...
Page 148: ...147 Safety and Regulatory Information 5 Safety and Regulatory Information ...