38
Chapter 2
Installation
Preparing the 11974 for Use
Insta
lla
tion
Using a Conversion-Loss Data Disk with the ESA or PSA Series
Analyzers
The conversion-loss data for your mixer can be quickly loaded in the memory of an ESA or
PSA series analyzer from a floppy disk. The following process works with the floppy disk
provided by the factory with your new mixer, or with one that you have stored the
correction data. This process eliminates the time consuming and sometimes error-prone
process of manually entering the data from a table.
Loading Data
1. Insert the conversion-loss data disk (from the factory or your own) into the floppy drive
of the analyzer.
2. Press
File
, then
Load
.
3. Press
Type
,
More
, then select
Corrections
.
4. Press
Directory Select
, highlight drive
-A-
, then press
Directory Select
again.
5. Use the down arrow to select the file. The factory supplied disk has a file with the
OTH
extension which places the data in the corrections set “other” location.
6. Press
Load Now
to load the conversion loss values into the correction table of the
analyzer.
Viewing Data
1. Press
Amplitude
,
More
,
Corrections
,
Other
(ON) and
Edit
to view the correction values.
2. Use the front-panel knob or up/down arrows to scroll through the list of correction
values.
Summary of Contents for 11974 Series
Page 5: ...4 ...
Page 8: ...7 General Information 1 General Information ...
Page 28: ...27 Installation 2 Installation ...
Page 47: ...46 Chapter2 Installation Returning Equipment for Service Installation ...
Page 48: ...47 Operation 3 Operation ...
Page 97: ...96 Chapter3 Operation Using 8561A 62A 62B Spectrum Analyzers Operation ...
Page 98: ...97 Performance Tests 4 Performance Tests ...
Page 148: ...147 Safety and Regulatory Information 5 Safety and Regulatory Information ...