Chapter 4
111
Performance Tests
1. Conversion Loss
P
e
rf
orm
a
nce
T
est
s
15.Set the spectrum analyzer controls as follows:
SHIFT
,
5
,
MHz
,
0
, and
dB
(sets external mixer conversion loss to 0 dB)
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 MHz
ENTER dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 dB
CF STEP SIZE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 MHz
16.Set the reference level of the spectrum analyzer by pressing
REFERENCE LEVEL
and
entering the corresponding value for the appropriate band listed below:
for an 11974A:
−
40 dBm
for an 11974Q:
−
40 dBm
for an 11974U:
−
40 dBm
for an 11974V:
−
55 dBm
17.Set the center frequency of the spectrum analyzer by pressing
CENTER FREQUENCY
and
entering the corresponding value for the appropriate band listed below:
for an 11974A: 26.5 GHz
for an 11974Q: 33 GHz
for an 11974U: 40 GHz
for an 11974V: 50 GHz
18.On the spectrum analyzer, press MARKER
PEAK SEARCH
and
PRESEL PEAK
.
19.Record the following values in
Marker Frequency
Marker Amplitude
Power Meter Reading
Cal Factor % or Correction Factor (dB)
Directional Coupler Coupling Factor
NOTE
For the purpose of this measurement, the directional coupler coupling factor
is defined as the ratio of the power at the output flange to the power at the
coupled flange.
Summary of Contents for 11974 Series
Page 5: ...4 ...
Page 8: ...7 General Information 1 General Information ...
Page 28: ...27 Installation 2 Installation ...
Page 47: ...46 Chapter2 Installation Returning Equipment for Service Installation ...
Page 48: ...47 Operation 3 Operation ...
Page 97: ...96 Chapter3 Operation Using 8561A 62A 62B Spectrum Analyzers Operation ...
Page 98: ...97 Performance Tests 4 Performance Tests ...
Page 148: ...147 Safety and Regulatory Information 5 Safety and Regulatory Information ...