56
W-Class
Standard Calibration (continued)
Step
Operator Action
Displayed Message
Comment
3
Press the
key to
enter the Media Settings
menu.
!
NFEJB!TFUUJOHT!
NFEJB!UZQF!
!!
!
!
See Section 4.1.2
for menu layout.
4
Press the key
and scroll to ‘Sensor
Calibration’.
!
NFEJB!TFUUJOHT!
TFOTPS!DBMJCSBUJPO!
!!
!
Press the
key to abort this
procedure.
5
Press the
key.
!
TFOTPS!DBMJCSBUJPO!
QFSGPSN!DBMJCSBUJPO!
!
You are beginning
the standard
calibration
procedure.
6
Press the
key.
Remove the media from
the Media Sensor then
press any key.
!!
SFNPWF!MBCFM!TUPDL!
QSFTT!BOZ!LFZ!=zzz?!
!
This sets the
parameter for the
‘empty’ value.
Where ‘yyy’ is a
numerical value
representing the
current sensor
reading.
Summary of Contents for W Class
Page 2: ......
Page 4: ......
Page 10: ...iv Warranty Information 113 Glossary 117 Index 121 ...
Page 32: ...22 W Class ...
Page 96: ...86 W Class ...
Page 120: ...110 W Class ...
Page 126: ...116 W Class ...