Vibration
5-200 Hz (vertical, longitudinal, transversal
axis)
Environmental conditions for mobile installation (during transport)
In line with IEC721-3-5: 5K2 and 5M2 with following restrictions:
Vibration
5-150 Hz (all axis), 1m/s², sinusoidal vibration
Environmental conditions for mobile installation (during operation)
In line with IEC721-3-3: 3K2 and 3M1 with following restrictions:
Temperature
+15 °C to +30 °C
Relative Humidity
15% to 75% (non-condensing)
Vibration
40-200 Hz; 1m/s²; sinusoidal vibration
Warming-up time
Cold start
fully operational after max. 30 min.
Warm start
fully operational after self-test, on condition
that the digitizer:
• has not been switched off for more than
3 minutes.
• has been operational for at least 30 minutes.
Physical emissions
Noise emission (sound power level according to ISO 7779)
• During scanning
max. 65 dB(A)
• Standby
max. 55 dB(A)
Heat emission
• Standby
60 W
≈
204 BTU/h
1
• Average power consumption during
scanning
CR 30-X: 85 W
≈
290 BTU/h
1
CR 30-Xm: 103 W
≈
351 BTU/h
1
• Peak power consumption during scanning
CR 30-X: 190 W
≈
648 BTU/h
1
CR 30Xm: 220 W
≈
751 BTU/h
1
End of Life
Estimated product life
(if regularly serviced and maintained
according to Agfa instructions)
7 yrs.