60
2386G EN 20120110
Remarks for HF-emission and immunity
CR 30-X/CR 30-X
M
Remarks for HF-emission and
immunity
The device has been tested for a normal hospital environment as described
below.
The user of the device should ensure that it is used in such an environment.
Nevertheless the HF-emission and immunity can be influenced by connected
data cables depending on length and the manner of installation.
Emissions test
Compliance
Electromagnetic Environment
Guidelines
RF emissions in accordance with
CISPR 11
Group 1
The digitizer uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions in accordance with
CISPR 11
Class A
The digitizer is suitable for use in all
establishments other than domestic, and
may be used in domestic establishments
and those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes, provided the following warning
is heeded.
WARNING
: This device is intended for use
by healthcare professionals only. This
device may cause radio interference or may
disrupt the operation of nearby equipment.
It may be necessary to take mitigation
measures, such as re-orienting or
relocating the digitizer or shielding the
location.
Harmonic emissions in
accordance with IEC 61000-3-2
Class A
Voltage fluctuations / flicker
emissions in accordance with
IEC 61000-3-3
Complies