55
2386G EN 20120110
Equipment Information Sheet
CR 30-X/CR 30-X
M
Power consumption
Standby
• 220 V - 240 V / 50-60 Hz configuration
60 W
• 100 V - 120 V / 50-60 Hz configuration
60 W
During operation
• 220 V - 240 V / 50-60 Hz configuration
CR 30-X: max. 190 W
CR 30-Xm: max. 220 W
• 100 V - 120 V / 50-60 Hz configuration
CR 30-X: max. 190 W
CR 30-Xm: max. 220 W
Uninterruptible power supply (optional)
UPS Powerware 5115
120 V
ABC ordering code: EGPSE
UPS Powerware 5115
230 V
ABC ordering code: EGPTG
Environmental conditions
Room temperature
recommended: 20 °C to 25 °C
allowed: 15 °C to 30 °C
Maximum temperature change
0.5 °C/min.
Relative humidity
recommended: 30 % to 60 %
allowed: 15 % to 75 % (non-condensing)
Magnetic field
compliant with EN 61000-4-8, Level 2
Sunlight exposure
not to be operated in direct sunlight, max.
2500 lux
Environmental conditions (during storage)
In line with IEC721-3-1: 1K2 (CR 30-X) / 1K4 (CR 30-Xm) and 1M2
Temperature
-25 °C to +55 °C
Environmental conditions (during transport)
In line with IEC721-3-2: 2K2 and 2M2 with following restrictions:
Temperature
-25 °C to +60 °C