Introduction
CR 30-X/CR 30-X
M
Mobile Use Installation
In case of an installation in a mobile environment, such as a bus, van, etc, the
manufacturer of the vehicle should ensure that all components of the system
are fixed or can be fixed safely for transport.
If the digitizer is installed in a mobile environment, it has to be secured
against moving. The optional earthquake kit for wall fixation should be used.
Image quality check after transport
The image quality check must be performed after installation of the digitizer in
a mobile environment and is recommended to be repeated after transport.
The check is done with a flat field exposure and should be performed with a
cassette of the largest format used at customer site.
WARNING:
The device is a table-top digitizer. The structure and
stability of the table used, need to be suitable in relation with the
size and weight of the system. The table should not be subject to
excessive shock and vibrations from other sources, as this may
disturb the operation of the digitizer.
X-Ray source
Exposure conditions
General radiography
It is recommended to expose the cassette with 2 exposures of
10 µGy or 1 mR each. Rotate the cassette 180° after the first
exposure to compensate for the heel effect.
Typical settings for 10 µGy or 1 mR are:
• 75 kV
• 12 mAs
• 130 cm SID
• large focus
• 1.5 mm Copper filter
Identify the cassette as System Diagnosis GenRad - Flat Field.