TROUBLE
PROBABLE CAUSES
POSSIBLE SOLUTIONS
Fault message E 011 appears
Hardware problems in the power supply.
Turn the equipment off and then on
again. If the trouble persists, replace
the main PC board
Fault message E012 appears
Improper operation
Press any key and then perform the
operation in the correct manner
Fault message E 013 appears
The audible warning device does not work
Replace the main PC board
The control circuit of the audible warning
Replace the main PC board
device does not work
Fault message E 014 appears
Hardware problem in the control circuit
Turn the equipment off and then on again
If the trouble persists, replace the main PC
board
Fault message E 015 appears
Hardware problem in the voltage tester
Turn the equipment off and then on again
circuit
If the trouble persists, replace the main PC
board
Fault message E EKE appears. There
One or more conductors are interrupted
Replace the cable
is a communication problem between
Hardware problems in the control panel
Replace the handpiece control assembly
the main PC board and the control
Hardware problems with the main PC
Replace the main PC board
panel
board
Film too light
Connection wires to the tubehead are
Check wiring
reversed
The value selected by the “F” key is
Press the “F” key to raise the value
too low for the chosen film
The temperature of the developer and
Check the temperature of the agents or
fixing agent is incorrect or they are in poor
replace them
condition
X‑rays and audible warning device are
Tubehead faulty
Replace the tubehead
not emitted when the X‑ray control
button is pressed
If the READY light is off, this means the
Wait for the light to come on
tubehead is cooling down
A wire inside the connecting cable
Make sure the connectors are properly
between the control panel and main PC
plugged in, if necessary replace the
board is interrupted
coil‑cable
Hardware problem on the X‑ray control
Replace the main PC board
circuit of the control unit
Fuse F1 in the control unit has blown
Replace the fuse with one having the same
amperage rating, F10 A ‑ 250V (5x20)
One or more lights on the control panel
Lights have burned out
Replace the handpiece control assembly
have not come on.
During X‑ray emission the audible
Audible warning device is faulty
Replace the main PC board
signal is not given out
Timer makes clicking sound with
No connection to the handpiece
Check the connection
no display
control assembly
Replace the handpiece control assembly