B4. Corrective maintenance
B4.1. Service test
CAUTION: TO BE PERFORMED ONLY BY A QUALIFIED SERVICE
TECHNICIAN
Press and hold the “TEST” button (e) while turning on the system to enable the “Service Test” mode.
The display (a) indicates the type of test pro cedure that has been selected - refer to paragraph
A2. Use the Test button (e) to change over to the next test.
TEST 0
- EPROM memory version in the control panel. The display (b) indicates the EPROM
memory version in the control panel.
TEST 1
- EPROM memory version in the control unit and check out procedure on the indicator
lights and pushbuttons. The display (b) indicates the EPROM memory version in the control unit.
Moreover, this test is used to check the pushbuttons and indicator lights located on the control
panel. To do this, proceed as follows:
· Press any button in order to turn on the related light. The “X-ray” and “Ready” lights (g & f) will
come on and off one after the other every time a button is pressed.
· Press the “Test” button (e) to change over to the next test.
TEST 2
- Function enabling
This test is used to check the state of the jumpers on the control unit card.
· Press the “F” button (d) to select the number of the jumper shown on the display. The letter on
the left indicates the state of the jumper: A= open; C = closed
TEST 3
- Checking the connections
This test is used to check the inputs and outputs between the control unit and the X-ray tubeheads
installed.
· Bring one tubehead at a time into its work position.
· The LED on the wall-mount plate lights up.
· The display (b) indicates the wall mount plate
selected (a, b, c, d).
· Press the X-ray control button (h).
If “P - E” is shown on the display’ the
command is okay.
TEST 4
Supply voltage
During this test procedure’ the control panel
serves as a voltmeter and the display (b) will
show the input voltage to the equipment.
The voltmeter is set by the manufacturer
with high precision instruments during the
testing procedures. If a difference greater
than 3 V is found between the voltage shown
on the display and the rated voltage, the
equipment must be reset as next described.
d
e
f
a
b
j
i
h
g