17
EXPOSE
S
NOTE: All upper and lower tooth selections are for
dogs only. Due to the relatively small dentition of cats,
only the selection for any upper or maxillar feline tooth
(CAT UPPER) or any lower or mandibular feline tooth
(CAT LOWER) is needed. The PAW setting is for any
paw that can be rediographed within the field size of
the cone. The EXOTIC setting is for bird wings, small
animal extremities, etc., that can be radiographed
within the field size of the positioning cone.
The above chart shows the amount of exposure
time variation for the different film sensitivity values
Radiograph-film correction chart
Sensitivity
Standard Variation
coefficient
1
0.500
- 50 %
2
0.630
- 37 %
3
0.800
- 20 %
4
1.000
0%
5
1.250
+ 25 %
6
1.600
+ 60 %
7
2.000
+ 100 %
8
2.500
+ 150%
9
3.200
+220%