B2.1.1. X-Mind Timer Assembly direct or on optional Wall Mount Plate
First, determine if the X-Mind Timer Assembly will be mounted directly on the wall or if site condi-
tions require the wall mount plate. This will be determined by the existing wall construction and
load bearing ability. The wall mount plate spreads the load over a larger area and is recommended
for dry wall/stud construction sites.
1. To be sure that the equipment is in the correct position, we recommend you put the
provided template (
3
) into the desired position. Considering the overall dimensions of the
equipment, put the top part of the template approximately 57” from the floor.
2. Mark the mounting points and make the respective holes with a diameter corresponding to
the chosen screws, depending on the wall construction in the room.
3. Remove the plastic timer cover (
1
) by loosening the two screws (2)located at the bottom
and lifting the cover from the bottom to the top to release the upper clamps.
4. Attach the timer to the wall using the relevant screws (
4
).
Template
X-Mind timer/Wall mount
with circuit board (hand-held
control not shown)
Front cover