Zeus Hand Configurator - Web Application Instructions
27
Prosthesis Settings Tab
12
Hold-open - hold-open timings. In these settings there are two sliders
used for specifying the time for primary change signal and secondary
change signal.
Soft-grip
The strength of the fingers can be limited by the soft-grip functionality. It adjusts
the maximum grip force applied by the hand on an object. It also enables you to
specify an additional threshold for the EMG closing signal, which once exceeded,
restores the full strength. By default then, the patient can have a very light touch,
but likewise can squeeze and even squash an object if desired. When enabling this
function, you need to calibrate the fingers by pressing the Calibration procedure
button.
WARNING
: Keep hand in a safe position while calibrating as its fingers will be
moving.