Zeus Hand Configurator - Web Application Instructions
25
In hold-open mode, the patient generates the Secondary Change Signal
by keeping the open signal high while the fingers are already fully open
for 1.5 seconds and he generates the Secondary Change Signal by
doing the same for 2.5 seconds. These timings can be adjusted by
changing the Hold-open grip switching mode sliders.
Single electrode grip switching mode is used when only one input / signal
is possible. In this mode, the patient closes and opens the hand with the
same signal. The first generation of the signal closes the fingers, the
second generation of the signal opens the fingers. Double short impulse
of the signal is used for a primary grip change signal and triple short
impulses are used for a secondary grip change signal.
Prosthesis Settings Tab
12
Grip switching modes
These settings will appear when correct grip switching mode is selected
Co-contraction - Co-contraction timings. Long co-contraction time allows
you to specify the time after which holding a co-contraction signal is
treated as a secondary grip change signal. A signal shorter than this value
is treated as the primary grip change signal. The signal rise offset time
slider allows you to specify the maximum time between two signals
crossing the Change Signal (CS) threshold to be treated as a grip change
signal.