Zeus Hand van Aether Biomedical, distributeur Loth Fabenim B.V.
EN
Zeus Hand Configurator - Web Application Instructions
23
Prosthesis Settings Tab
12
In the Prosthesis settings tab you are able to configure various settings and
parameters for the Zeus hand.
Input Options
This refers to the method by which a patient can control the prosthesis namely
via EMG electrodes that are accepted with the Zeus hand.
Input Sites
The choice of this option depends on the patient’s muscle condition, the level
of the amputation and their stage in the rehabilitation process. It relates to the
production and sending of good quality EMG signals.
Dual Direct
Dual Inversed
Single
These options mean the patient can use one electrode or two - in direct or in
versed mode - to control the prosthesis.
Dual direct mode is the default option, with signals from two electrodes being
used to control the prosthesis.
Dual inverted mode simply swaps the electrode inputs. The electrode that would
close the prosthesis’ fingers in the direct mode will open them in the inverted
mode.
Single electrode mode is designed for patients that cannot use dual modes
because of difficulty in producing two differentiable, strong signals. When
selected, the application will automatically choose “single electrode” grip
switching mode.