Zeus Hand Configurator - Web Application Instructions
18
EMG Settings Tab
10
In the EMG Setting tab you can visualize the graph depicting the signals and their
strength acquired from the patient’s muscles.
To accommodate for varying levels of signal strength and electrode sensitivity,
there are special slider adjustments to change the threshold levels for the grip
change signal and the activation levels to open and close the hand.
On the left side of the screen there are 2 sliders with a scale. They can be easily
adjusted to establish a suitable grip change signal threshold for each patient
individually.
On the right side of the screen there are 2 additional sliders. They can be used
to adjust a suitable activation level for the open and close signal.
You can adjust the timescale on the graph by slider at the bottom of Settings
Area.
You can hide or show the thresholds on the graph by toggling the checkbox
beneath their sliders.
You can adjust the gain of the electrodes to change the strength of the signals
by moving the sliders on the bottom of the screen. It is recommended to change
the gains on the electrodes first, and adjust the software gains for fine-tuning.
There are a couple of options that affect the appearance of the graphs.
Combined - in this option, both muscle signals (Flexion and Extension)
are shown on the same graph