Zeus Hand Configurator - Web Application Instructions
22
Grip Configuration Tab
11
In the Grips configuration tab you can change the initial positions and position
limits in any grip.
The initial position of each finger is the position that is assumed after changing
to the grip. Position limit is the maximum position the finger can move to while
closing. Both of those parameters can be adjusted for each finger individually in
every grip; 1000 is fully closed, 0 fully open. This can be used for fine tuning of
the grips.
To begin, you have to select the right grip in the dropbox at the top right corner.
You can activate the movement of the hand while configuring the grips to see
what the finger positions actually look like by checking the box at the top right
corner.
WARNING
: Be wary of the surroundings of the hand while doing this try to avoid
moving fingers in a way that they might hit other digits, as this may lead to damaging
the fingers.